Method
Min. waiting time
[ 0 ] ... 999'999 s
The measured value is only accepted if the minimum waiting time has
elapsed even if the signal drift has already been reached. The drift contin-
ues to be checked as the minimum waiting time is passing.
Max. waiting time
0 ... [ 52 ] ... 999'999 s
If the signal drift has not yet been reached then the measured value will be
accepted when the maximum waiting time has elapsed. If no new entry for
the waiting time has been made then a suitable value for the drift will be
calculated automatically according to the following equation:
Measuring interval
0.1 ... [ 2.0 ] ... 999'999.0 s
Time interval for entering a measuring point in the list of measuring points.
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150 °C
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement
on the tab
General/Hardware
is set to
automatic
or
continuous
, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.
Switch off stirrer during measurement
on, [ off ]
If this option is switched on then the stirrer is switched off during the meas-
urement. In this case the two following parameters can be edited:
Stir before measurement
[ 0 ] ... 999999 s
Stirring time between end of dosing and measurement.
Pause before measurement
[ 0 ] ... 999999 s
Waiting time before the measurement during which no stirring takes place.
STDADD dos
Command for Standard addition with addition of the standard addition solution
from a dosing device.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
STDADD dos
are configured on the following 3
tabs:
•
General/Hardware
Parameters for device, dosing device, sensor and stirrer.