Workplace

Modification comment for sample data

If the option

Comments on modification of sample data

is switched on in the Security

settings then before the edited sample data is accepted the window

Modification

comment sample data

opens first, in which a reason for the modification must be

selected and comments about the modification must be entered.

Reason

Selection from standard texts

Slection from the Default reasons defined in the dialog window

Security set-

tings

for the category

Modification of sample data

.

Comment

1000 characters

Entry of comments about the modification to the sample data.

Determination run

A determination that is started on the tab

Single determination

in the subwindow

Run

takes place as follows:

1 - Load sample data

The Sample data entered on the tab

Single determination

is loaded for the deter-

mination.

2 - Select method

Method selection triggers the following actions:

Select method without assignment ID

If the option

Method selection by assignment ID

is not switched on on

the tab

Process

then the method selected in the field

Method

will remain

loaded.

Select method with assignment-ID

If the option

Method selection by assignment ID

is switched on on the

tab

Process

then the

Sample assignment ID

will be used to load the

method defined in the

Sample assignment table

.

3 - Carry out start test

During the Start test the following checks and actions are triggered:

Right to carry out the method?

Checks whether the logged-in user has the right to execute the par-
ticular method.

Method executable?

Checks whether the particular method is executable. If the method is
non-executable then a message appears asking for the Method test to
be carried out and the method edited accordingly.

Check device (module) and rack

Checks whether the devices (or device types), modules (e.g. measur-
ing inputs, towers, etc.) and racks demanded in the device-specific
commands are present and available or assigned. If the device has