beautypg.com

Metrohm tiamo 1.1 Manual User Manual

Page 727

background image

Subwindow Devices

699

COM Port

COM1, COM2, ..., [ first free COM port ]

Selection of the serial interface on the PC to which the device is connected.

Baud rate

1200, 2400, 4800, [ 9600 ]

Transmission speed. Additionally, this baud rate must be set on the device
itself.

Note

Thes parameters are only editable for devices with status

not ok

(power and/or

RS connection interrupted).

GLP

GLP test date

Date selection

Date of the last GLP test. This date can be selected after pressing

in the

dialog window

Select date

(see General program functions - Edit - Select

date).

Comment on GLP test

1000 characters, [ empty ]

Comments about GLP test.

Monitoring of GLP validity

on, [ off ]

If this option is switched on then the time interval for the GLP test will be
monitored.

Note

This option can only be switched on when a date has been entered under

GLP

test date

.

GLP test interval

0 ... [ 999 ] days

Time interval to next GLP test. If an interval is entered here then the date for
the

Next GLP test

will be adapted automatically.

Next GLP test

Date selection, [GLP test date + 999 days ]

Date on which the next GLP test is to be carried out. This date can be se-
lected after pressing

in the dialog window

Select date

(see General

program functions - Edit - Select date). After the date has been entered the
value for the

GLP test interval

will be adapted automatically.