Titration parameters – Metrohm tiamo 1.1 Manual User Manual
Page 390

362
Method
Titration parameters
Parameters defining the run of the titration.
Command name
25 characters, [ DET Upol # ]
Name of the command.
Titration rate
Titration rate
Three predefined sets of parameters can be selected for the titration rate:
slow
,
optimal
and
fast
. For these sets the parameters for the
Measured value
acceptance
and
Dosing of increments
are not displayed. In order to edit these
parameters the value
user
must be selected.
slow
For titrations in which the finest details are also to be visible. However, this
can also increase noise and lead to the production of unwanted EPs.
[ optimal ]
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical rapid titrations.
user
Editing the individual titration parameters which affect the titration rate.
Measured value acceptance
(only visible for
Titration rate
=
user
)
Signal drift
0.01 ... [ 20.00 (slow) ] ... [ 50.00 (optimal) ] ... [ 80.00 (fast) ] ... 99.90 µA/min
(Titrando)
0.05 ... [ 20.00 (slow) ] ... [ 50.00 (optimal) ] ... [ 80.00 (fast) ] ... 99.90 µA/min
(Titrino)
off
Drift for accepting the measured value.
off
means that the measured value
will be accepted when the maximum waiting time has elapsed.