beautypg.com

Data import – Metrohm tiamo 1.1 Manual User Manual

Page 151

background image

Subwindow Run

123

Autorepeat determination

on, [ off ]

If this option is switched on then at the end of a determination the next de-

termination with the same method will be started automatically. In order to
stop the autorepeat the determination has to be stopped with

[Stop]

.

Data import

Switches the import of external data on and off for the fields on the tab

Single de-

termination

.

Import from balance

on, [ off ]

If this option is switched on then external data from the selected balance
will be imported into the fields on the tab

Single determination

. Data can be

imported into the fields

ID1...ID4

,

Sample size

and

Sample size unit

.

Note

In order to be able to import sample identifications from the balance the data
export must be set accordingly on the balance (use names

ID1

,

ID2

,

ID3

,

ID4

).

Balance

Device name

Selection of the balance from which data is to be imported.

Always accept data

[ on ], off

If this option is selected then the data transmitted from the balance will be

imported into the tabs

Single determination

of all opened workplaces for

which this balance has been defined as the data source.

Accept data only for active window

on, [ off ]

If this option is selected then the data transmitted from the balance will only
be imported into the tab

Single determination

of the active workplace.

Import from barcode reader

on, [ off ]

If this option is selected then external data from the selected barcode
reader will be imported in the fields on the tab

Single determination

. The field

into which the data is to be imported is defined in the device properties of
the Barcode reader.

Barcode reader

Device name

Selection of the barcode reader from which data is to be imported.