Evaluations – Metrohm tiamo 1.1 Manual User Manual
Page 490

462
Method
Stop measured value
-2'000.0 ... 2'000.0 mV, [ off ]
(Titrando only)
Stop when the preset measured value has been reached since the start of
the measurement.
off
means no stop.
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500 °C
(Titrino)
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement
on the tab General/Hardware is set to
automatic
or
continuous
, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.
Evaluations
The following methods for evaluating measuring curves are defined on this tab.
Command name
25 characters, [ MEAS U # ]
Name of the command.
Fix end point evaluation
on, [ off ]
To a fixed quantity (measured value or time) the associated values for the
other quantities are interpolated from the list of measuring points for the
fixed end point.
Fixed quantity
[ Measured value ], Time
Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measuring points.
Fix EP# at
-2'000.0 ... 2'000.0 mV, [ off ]
(Titrando)
-2'000 ... 2'000 mV, [ off ]
(Titrino)
Fixed end point 1 ... 9 for
Fixed quantity
=
Measured value
.