Signatures – Metrohm tiamo 1.1 Manual User Manual
Page 629

Administration
601
Audit Trail
Enable Audit Trail
on, [ off ];
FDA-default:
on
If this option is switched on then all program actions will be automatically
recorded; these are defined as Audit Trail actions.
Modifications
Comment on modification of methods
on, [ off ];
FDA-default:
on
If this option is switched on then each time that a method is altered a
modification reason and a modification comment must be entered; these
are saved in the method and shown in the method history. The reason and
comments are also recorded in the Audit Trail.
Comment on modification of determinations
on, [ off ];
FDA-default:
on
If this option is switched on then each time that a determination is altered
a modification reason and a modification comment must be entered; these
are saved in the determination and shown in the database in the subwin-
dow
Information
on the tab
Determination
. The reason and comment are
also recorded in the Audit Trail.
Comment on modification of sample data (live)
on, [ off ];
FDA-default:
on
If this option is switched on then each time that sample data is altered an
modification reason and modification comment must be entered; these are
saved in the determination and shown in the database in the subwindow
Information
on the tab
Sample
. The reason and comment are also recorded
in the Audit Trail.
Signatures
On the tab
Signatures
the parameters for the electronic signatures can be config-
ured.