Cal loop ph – Metrohm tiamo 1.1 Manual User Manual
Page 529

Commands
501
under the parameter
Target
:
Move
:
Special beaker
Number
:
= ### + 'Command name.LCO'
(entry with Formula editor, i.e.
= 5 + 'Calibration loop pH.LCO'
for the above sample method with the
first calibration solution on special beaker position
6
).
• Defining rack positions for calibration solutions in the sample
data
A
Sample position
for the first calibration solution has to be defined in
the run window. The other calibration solutions must be placed on the
directly following rack positions. The following settings have to be
done in the command
MOVE
under the parameter
Target
:
Move
:
Rack position
Number
:
= 'MV.Sample position' – 1 + 'Command name.LCO'
(entry with
Formula editor, i.e.
= 'MV.Sample position' – 1 + 'Calibration loop
pH.LCO'
for the above sample method).
The variable
'Command name.LCO'
(index number of the calibration loop) is in-
creased by
+1
after each pass of the loop.
Note
The Formula editor is opened with a right click in the input field.
CAL LOOP pH
Loop command for calibrating pH electrodes. The command is used together
with the command
CAL MEAS pH
.
Appearance
The command has the following appearance:
Parameters
The parameters for the command
CAL LOOP pH
are configured in the following dia-
log window:
•
CAL LOOP pH