Subwindow Rack data
757
Note
Special beakers should preferably be set at high rack positions, so that a sam-
ple series can start at rack position 1. Rack positions that are defined as special
beakers are no longer available as sample positions. A specific work position
can be defined for each special beaker at Tower 1 and Tower 2 (if present). The
shift, rinse and special positions for the particular tower are taken over from the
general rack positions.
The following columns are shown in the table:
Special beaker
Number of the special beaker for the selected rack.
Rack position
Number of the rack position for the special beaker.
Work position Tower 1
Work position for the special beaker at Tower 1.
Work position Tower 2
Work position for the special beaker at Tower 2.
Beaker radius
Radius of the special beaker.
Beaker sensor
Shows whether and which beaker sensor is to be used for the special
beaker.
Opens the dialog window
Special beaker #
for editing the data of the se-
lected special beaker.
Special beaker
In the dialog window
Special beaker #
the parameters for the special beaker se-
lected in the Table of special beakers can be defined.
Rack position
[ 0 ] ... n (n is rack-dependent)
Number of rack position for selected special beaker.
Note
Special beakers should preferably be set at high rack positions, so that a sam-
ple series can start at rack position 1. Rack positions that are defined as special
beakers are no longer available as sample positions. They will be jumped over
during an automatic movement to a sample position.
Work position Tower 1
[ 0 ] ... 235 mm
Work position of the selected special beaker at Tower 1.
Work position Tower 2
[ 0 ] ... 235 mm
Work position of the selected special beaker at Tower 2.