Column display, Add new solution – Metrohm tiamo 1.1 Manual User Manual
Page 743

Subwindow Titrants/Solutions
715
Column display
With
Edit, Column display...
the dialog window
Column display
opens. Here you can
define the columns to be shown in the Solution table.
Columns available
Shows all the fields that can be shown as columns in the solution table.
Columns displayed
Shows all the fields that will be shown as columns in the solution table. The
default situation is that the columns
Solution name, Concentration, Cylinder
volume, Type, Dosing device, Titer, Date titer det., Next titer determination
and
Expiry date
are shown. The column
Solution name
is always present and
cannot be removed.
Adds the selected column to the table.
Removes the selected column from the table.
Alters the sequence of the shown columns by moving the selected column
upward or downward.
Add new solution
Solutions in non-intelligent exchange units (EU) and dosing units (DU) without
data chip must always be added manually to the solution table with
Edit, New, Ex-
change unit
or
Edit, New, Dosing unit
. This automatically opens the Properties win-
dow for configuring the solution. After the Properties window has been closed the
solution with its set parameters will be entered in the solution table. The parame-
ters can be altered at any time with
Edit, Properties
.
Note
Solutions in intelligent 806 Exchange Units (IEU) or 807 Dosing Units (IDU) with
data chip must only be manually added and configured when they are attached
to devices that cannot read in the data automatically (e.g. Titrino, 700 Dosino).