Additional evaluations – Metrohm tiamo 1.1 Manual User Manual
Page 446

418
Method
Drift correction
automatic, manual, [ off ]
If this parameter is set to
automatic
or
manual
the drift correction ( =
Drift
value
x Titration time) caused by blank consumption during titration is de-
ducted from the total consumption. With
automatic
the drift value at the start
of the titration applies with
manual
the value defined under
Drift value
is ap-
plied.
Drift value
[ 0 ] ... 99.9 µL/min
Drift value for drift correction. This parameter is only shown for
Drift correc-
tion
=
manual
.
Delay after 'Cond ok'
[ 0 ] ... 999 s
Waiting period until the start of the titration during which the conditioning
condition must be fulfilled.
Only start titration by a start command from a SEND command
on, [ off ]
If this option is set the titration is only started when the
Event message
Start
titration
from a
SEND
command reaches this command.
Note
At the time of reception of the
SEND
command conditioning must already be
terminated (=
Condok
) otherwise the command to start the titration will be ig-
nored.
Show measured value during conditioning
on, [ off ]
If this option is set additionally to the drift and time the measured value is
shown in the Live display.
Additional evaluations
The following additional method for evaluation of titration curves can be activated
and defined here:
Command name
25 characters, [ KFC # ]
Name of the command.
Fix end point evaluation
on, [ off ]
For a fixed measured value for the fix end point the associated values are
interpolated from the list of measuring points.