Drift correction

automatic, manual, [ off ]

If this parameter is set to

automatic

or

manual

the drift correction ( =

Drift

value

x Titration time) caused by blank consumption during titration is de-

ducted from the total consumption. With

automatic

the drift value at the start

of the titration applies with

manual

the value defined under

Drift value

is ap-

plied.

Drift value

[ 0 ] ... 99.9 µL/min

Drift value for drift correction. This parameter is only shown for

Drift correc-

tion

=

manual

.

Delay after 'Cond ok'

[ 0 ] ... 999 s

Waiting period until the start of the titration during which the conditioning
condition must be fulfilled.

Only start titration by a start command from a SEND command

on, [ off ]

If this option is set the titration is only started when the

Event message

Start

titration

from a

SEND

command reaches this command.

Note

At the time of reception of the

SEND

command conditioning must already be

terminated (=

Condok

) otherwise the command to start the titration will be ig-

nored.

Show measured value during conditioning

on, [ off ]

If this option is set additionally to the drift and time the measured value is
shown in the Live display.

Additional evaluations

The following additional method for evaluation of titration curves can be activated
and defined here:

Command name

25 characters, [ KFC # ]

Name of the command.

Fix end point evaluation

on, [ off ]

For a fixed measured value for the fix end point the associated values are
interpolated from the list of measuring points.