Rs 232 – Metrohm tiamo 1.1 Manual User Manual
Page 680

652
Configuration
Request for dosing device preparation
Selects when the request for carrying out a
PREP
(prepare) command is to be
shown for the external dosing device
D1
or
D2
.
At program start
[ on ], off
If this option is switched on then at each program start the request to pre-
pare the dosing device will appear.
Time interval
on, [ off ]
If this option is switched on then the request to prepare the dosing device
will appear after the time interval defined here.
all
0.1 ... [ 12.0 ] ... 999.9 h
Time interval after which the request to prepare the dosing device will ap-
pear.
Connected devices
The
Dosing device type
(685 Dosimat or 700 Dosino) connected to the dosing de-
vice connection is shown here if it has been defined in the device itself.
RS 232
COM Port
COM1, COM2, ..., [ first free COM port ]
Selection of the serial interface on the PC to which the device is connected.
Baud rate
1200, 2400, 4800, [ 9600 ]
Transmission speed. Additionally, this baud rate must be set on the device
itself.
Note
Thes parameters are only editable for devices with status
not ok
(power and/or