Prep - properties – Metrohm tiamo 1.1 Manual User Manual
Page 560

532
Method
PREP - properties
Command name
25 characters, [ Prepare # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Dosing device
Dosing device
[ 1 ] ... 4
(Titrando, 846)
[ 1 ] ... 3
(778, 789, 814, 815, 855)
[ 1 ] ... 12
(730, 774)
[ internal D0 ],
external D1, external D2
(736, 751, 758, 799)
Selection of the dosing device to be prepared. All the dosing device con-
nections (MSB) are displayed which are possible with the selected device
type.
Solution
24 characters, solution name, [ not defined ]
Input or selection of the solution defined in the solution table. Additionally,
not defined
can be selected. If intelligent exchange or dosing units are used
then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. At the start of the command the validity of the titer and
the expiry date of the selected solution will be checked as well as the GLP
test interval for the buret. If
not defined
is selected no tests will be carried
out.