Metrohm tiamo 1.1 Manual User Manual
Page 483

Commands
455
If the signal drift has not yet been reached then the measured value will be
accepted when the maximum waiting time has elapsed. If no new entry for
the waiting time has been made then a suitable value for the drift will be
calculated automatically according to the following equation:
Measuring interval
0.1 ... [ 2.0 ] ... 999'999.0 s
(Titrando, increment = 0.1 s)
0.08 ... [ 2.0 ] ... 16'200 s
(Titrino, increment = 0.08 s)
Time interval for entering a measuring point in the list of measuring points.
Stop measured value pH
-20.000 ... 20.000, [ off ]
(Titrando only)
Stops when the preset measured value has been reached since the start of
the measurement.
off
means no stop.
Measurement without drift control
[ on ], off
The measurement is carried out without drift control if this option is en-
abled. The measurement continues until one of the stop criterions
Measur-
ing time
or
Stop measured value pH
is met. The corresponding parameters
are not visible if this option is disabled.
Measuring time
0 ... [ 120 ] ... 999'999 s
(Titrando)
0 ... [ 120 ] ... 9'999 s
(Titrino)
Maximum period of time for measurement.
Measuring interval
0.1 ... [ 2.0 ] ... 999'999.0 s
(Titrando, increment = 0.1 s)
0.08 ... [ 2.0 ] ... 16'200 s
(Titrino, increment = 0.08 s)
Time interval for entering a measuring point in the list of measuring points.
Stop measured value pH
-20.000 ... 20.000, [ off ]
(Titrando only)
Stops when the preset measured value has been reached since the start of
the measurement.
off
means no stop.
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500 °C
(Titrino)
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement
on the tab General/Hardware is set to
automatic
or
continuous
, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.