Page 401: Additional measured values, Met u

Commands
373
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150.0 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500.0 °C
(Titrino)
Manually entered titration temperature. If a temperature sensor is con-
nected and the
Temperature measurement
in the tab
General / Hardware
un-
der
Sensor
is set to
automatic
or
continuous
then the temperature will be
measured continuously. This value is used for temperature correction in pH
measurements.
Stop conditions
Screenshot and parameters: see DET pH - Stop conditions
Potentiometric evaluation
Screenshot and parameters: see DET pH - Potentiometric evaluation
Additional evaluations
Screenshot and parameters: see DET pH - Additional evaluations
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
MET U
Command for Monotonic Equivalence Point Titrations with potentiometric
voltage measurement.
Devices
This command can be executed with the following devices which have the MET
mode:
Titrino: 702, 716, 721, 736, 751, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
MET U
are configured in the following 7 tabs:
•
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
•
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
•
Titration parameters
Parameters for the run of the titration.