9 automation commands, Move – Metrohm tiamo 1.1 Manual User Manual
Page 562

534
Method
Dosing device
Dosing device
[ 1 ] ... 4
(Titrando, 846)
[ 1 ] ... 3
(814, 815, 855)
Selection of the dosing device (Dosinos only) to be emptied. All the dosing
device connections (MSB) are displayed which are possible with the se-
lected device type.
Solution
24 characters, solution name, [ not defined ]
Input or selection of the solution defined in the solution table. Additionally,
not defined
can be selected. If intelligent exchange or dosing units are used
then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. At the start of the command the validity of the titer and
the expiry date of the selected solution will be checked as well as the GLP
test interval for the buret. If
not defined
is selected no tests will be carried
out.
5.6.9 Automation
commands
Commands for the operation of Sample Processors.
The following eight automation commands can be selected:
•
MOVE
Moving to a rack position or an external position.
•
SWING
Swinging of the robotic arm (with Swing Head only).
•
LIFT
Moving to a lift position.
•
PUMP
Switching on/off the connected or built-in pumps.
•
STIR
Controlling a connected stirrer.
•
RACK
Initialization of the rack attached.
•
HEATER
Controlling the oven temperature of the 774 Oven Sample Processor.
•
FLOW
Regulation of the gas flow of the 774 Oven Sample Processor.
MOVE
Command for moving to a rack position (sample position or special beaker).
Devices
This command can be executed with the following devices:
Sample Processor: 730, 774, 778, 789, 814, 815, 855