Metrohm tiamo 1.1 Manual User Manual
Page 158

130
Workplace
5 - Start method
At the start of the method the following checks and actions are triggered:
• Assign determination ID
An unambiguous determination ID is assigned for the determination.
• Reserve device (module)
The devices (or their functional units) used by the device-dependent
commands contained in the method are reserved for the duration of
the determination, i.e. they cannot be used in other workplaces nor
can they be operated manually.
• Increment sample number
The
Sample number
is increased by
+1
.
• Increment statistics counter
If statistics has been switched on in both the
START
command and on
the tab
Determination series
then the statistics counter will be increased
by
+1
. If a new method is loaded or if the statistics counter agrees with
the set counter then the current statistical data will first be deleted and
the statistics counter reset to
0
.
6 - Main run
When processing the loaded method the following actions can be carried out:
• Process series start track
At the start of the first determination in a series the Series start track is
started. If during the run of this track
[Stop]
is pressed then the Exit
track will start; if an error occurs the Error track starts.
• Process main track
After the start of the Main track the status changes to
BUSY
. The com-
mands of the main track and the other tracks that it calls up are proc-
essed in sequence. If the main track calls up tracks that are not free
then a pause will be made until these tracks can be started.
• Interrupt and continue run
A running determination can be interrupted at any time with
[Hold]
and
then continued with
[Cont]
. All the active tracks will be stopped and
continued together.
• Interrupt and continue series
A running series can be interrupted at any time with
[Pause]
. The de-
termination that is running in the series will be completed, but no new
determination will be started. With
[Cont]
the next determination in the
interrupted sample series will be started.
• Cancel sequence
A running determination can be canceled at any time with
[Stop]
. An
Exit track may be carried out; the determination will be terminated.
• Cancel by error
If when a determination is being carried out an error occurs that
causes the cancellation of the determination then an Error track may
be carried out; the determination will be terminated.
• Process exit track
When the main track and all the other tracks that is has called up have
been completed then an Exit track may be carried out; the determina-
tion will be terminated.
• Process series end track
At the end of the last determination in a series the Series end track will
start. If when this track is being processed
[Stop]
is pressed then the