Measuring parameters – Metrohm tiamo 1.1 Manual User Manual
Page 519

Commands
491
Addition mode
Fixed volume increments
[ on ], off
If this option is enabled a fixed volume increment muss be added for each
standard addition.
Addition volume
[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL
Quantity of the fixed volume increment.
Variable volume increments
on, [ off ]
If this option is enabled the defined volume increment muss be added for
each standard addition.
Addition volume 1...19
[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL
Quantity of the defined volume increments.
Addition solution
Concentration
0.001 ...[ 1.000 ] ... 999999.999
Concentration of addition solution.
Unit
mol/L, mmol/L, µmol/L, [ ppm ], g/L, mg/L, µg/L, mg/mL, µg/mL, %, mEq/L
Concentration unit of addition solution.
Measuring parameters
The measuring parameters are defined on this tab.
Command name
25 characters, [ STDADD man # ]
Name of the command.
Measurement with drift control
The measurement is stopped as soon as the
Signal drift
or the
Stop meas-
ured value
is reached or a
waiting time
has elapsed.
Signal drift
0.1 ... [ 10.0 ] ... 999.0 mV/min
The measured value is only accepted if the signal drift defined here has
been fallen short of.