4 signature level 2 – Metrohm tiamo 1.1 Manual User Manual
Page 45

Electronic signatures
17
User
Entry of the user name (short name).
Password
Entry of the password.
Reason
Selection from standard reasons
Selection from the Default reasons defined in the dialog window
Security
settings
for the category
Signature Level 1
.
Comment
1000 characters
Entry of remarks about the signature.
Signs for method or determination. The window remains open.
Note
Methods and determinations can only be signed at Level 1 when the user be-
longs to a user group that has the appropriate rights (see User administra-
tion/User group/Signatures).
2.3.4
Signature Level 2
In the
Signature Level 2
methods or determinations can be signed at Level 2.
Note
Methods or determinations that have been signed at Level 2 are
blocked, i.e.
they can neither be edited nor deleted. In order to be able to edit such methods
or determinations again the signatures at Level 2 must first be deleted (see De-
lete Level 2 signatures).
Info
Information about signatures and the deletion of signatures is shown in this
field. The following messages are possible:
Signature possible
The selected method or determination can be signed.
Signature 2 not possible (signature 1 missing)
The selected method or determination cannot be signed at Level 2 as it has
not yet been signed at Level 1.
Signature not possible (accessed by other client)
The selected method or determination cannot be signed as it has already
been marked for signature by another client.
User
Entry of the user name (short name).
Password