544
Method
RACK
Command for Initialization of the attached sample rack. The rack, lift and ro-
botic arm (if existing) are reset, the rack code is read off and the corresponding
rack data are transferred to the Sample Processor.
Devices
This command can be executed with the following devices:
Sample Processor: 730, 774, 778, 789, 814, 815, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
RACK
are configured in the following dialog win-
dow:
•
RACK
RACK - properties
Command name
25 characters, [ Initialize rack # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ 814 USB Sample Processor ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.