Request - sample data request – Metrohm tiamo 1.1 Manual User Manual
Page 605

Commands
577
REQUEST - sample data request
Message
Displaying the message defined in the command
REQUEST
.
Sample position
[ 1 ] ... 999
Position of the sample on the sample rack. This number can be used for
moving to a sample position with the command
MOVE
with the
Target posi-
tion
=
Sample position
.
ID1 ... ID8
-1.0E-99 ... 1.0E+99
(for
Type
=
Number
)
100 characters
(for
Type
=
Text
)
'YYYY-MM-DD'
or
'YYYY-MM-DD hh:mm:ss'
(for
Type
=
Date/Time
)
Sample identifications.
Note
In case the sample identifications are invalid at the start of the method because
they are entered with the
REQUEST
command during the run, the option
Check at
start
in the
START
command must be switched off for these method variables.
Sample size
[ 1 ], 10 numbers
Sample size.
Sample size unit
[ g ], mg, µg, mL,µL, pieces
Unit of the sample size (text only).
Note
Sample data can be read in from a
balance or a barcode reader. The data
import must therefore be switched on in the window Run (Single determination
or Determination series) and the corresponding devices must be defined. The
dialog window
Sample data request
will be closed automatically after the data
have been received from these devices.
In case the option
Comments on modification of sample data (live)
is activated in
the Security settings, the dialog window
Modification comment sample data
(see
Subwindow Run - Single determination - Modification comment sample data) is
displayed after entering the sample data. A
Reason
and a
Comment
must be en-
tered for the modification.