Additional measured values, Set u – Metrohm tiamo 1.1 Manual User Manual
Page 416

388
Method
With this button additionally to the always shown fix EP1...3 the fix EP4...9
are displayed.
With this button instead of fix EP1...9 only the fix EP1...3 are shown.
Minimum evaluation
on, [ off ]
For the minimum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation).
Threshold value
0.1 ... [ 1.0 ] ... 20.0 pH/mL
The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.
Maximum evaluation
on, [ off ]
For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation).
Threshold value
0.1 ... [ 1.0 ] ... 20.0 pH/mL
The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
SET U
Command for Set Endpoint Titrations with potentiometric voltage measurement.
Devices
This command can be executed with the following devices which have the SET
mode:
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 841, 842, 855
Appearance
The command has the following appearance:
without conditioning with conditioning