Commands
433
Action
Definition of the action to be taken when the limits of the temperature are in-
fringed:
Stop determination
(only Titrando)
The running
STAT
command is quit, then the exit track (if present) is started
and the determination is stopped.
Stop command
The running
STAT
command is quit, then the next command is started.
Wait for [Continue]
Reagent dosing in the current
STAT
command is interrupted and a message
appears. As soon as the monitored temperature is again within the limits
(including hysteresis) reagent dosing can be resumed by pressing the
[Continue]
button in this message window.
Wait for limit ok
Reagent dosing in the current
STAT
command is interrupted. As soon as the
monitored temperature is again within the limits (including hysteresis) re-
agent dosing is resumed automatically.
[ none ]
No action will be taken.
Track call on limit exceeding
In this table, which cannot be edited directly, max. 20 entries can be de-
fined for the track that is to be started automatically when a particular limit
is infringed.
Open the dialog window
Track call #
(see below) to enter parameters of a
new track call.
Open the dialog window
Track call #
(see below) to edit parameters of the
selected track call.
Delete the selected track call.
Track call
Monitoring
[ Measured value ], Dosing rate, Temperature, Any
Selects the quantity for which a track is to be started if its limits are infringed
(
Any
= any of the three quantities).
Limit exceeding
Lower limit, Upper limit, [ Any ], OK
Selects the limit infringement for which a track is to be started. With
Any
the
track will be started for an infringement of either the lower or the upper limit,
with OK the track will be started when the monitored quantity is again within
the limits (including hysteresis).