Dosing parameters – Metrohm tiamo 1.1 Manual User Manual
Page 544

516
Method
Temperature measurement
Type of temperature measuring (only Titrando):
continuous
A temperature sensor must be connected. The temperature is measured
continuously.
[ automatic ]
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered
Temperature
(tab
Titra-
tion parameters
) will used.
off
The temperature will not be measured. The
Temperature
(see Dosing pa-
rameters) which has been entered manually will be used.
Stirrer
Stirrer
[ 1 ] ... 4, off
(only Titrando)
Selection of the stirrer.
off
means that no stirrer will be used.
Stirring rate
-15 ... [ 8 ] ... 15
(only
Titrando)
Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.
Switch off automatically
[ on ], off
(only
Titrando)
The stirrer is automatically switched off at the end of the command if this
option is enabled.
Dosing parameters
The volume to be added, the dosing rate and the dosing time are defined on this
tab.
Command name
25 characters, [ DOS pH # ]
Name of the command.
Dosing criterion
[ Volume/Dosing rate ], Volume/Dosing time, Dosing rate/Dosing time
Selects the two parameters that can be preset for dosing. The third pa-
rameter is calculated automatically from the formula
Volume = dosing time *
dosing rate
and is not shownDOS - monitored dosing.
Volume
0.00000 ... [ 10.0000 ] ... 99999.9 mL
Fixed volume to be added. This parameters is only displayed for
Dosing cri-
terion =
Volume/Dosing rate
or