340
Method
Calculated measured value
With the option
Additional calculated measured values
3 new measured values can
be calculated from existing measured values. They can be displayed in curves
with the identification
Calculated 1...3
or used as variables
'Command name.CA1...3'
in formulae.
Name
25 characters, [ Calc. measured value # ]
Designation for the calculated measured value. This name is used as axis
legend.
Formula
1000 characters, [ 'empty' ]
Formula to calculate the measured value. It can either be entered directly or
can be created in the formula editor which opens with
.
Unit
16 characters, [ 'empty' ]
Unit of calculated measured value.
Decimal places
0 ... [ 2 ] ... 5
Number of decimal places with which the calculated measured value will be
displayed.
Comment
250 characters, [ 'empty' ]
Comment about the calculated measured value.
Additional external measured values
on, [ off ]
3 new measured values can be taken over from existing measuring com-
mands (i.e.
MEAS
) running simultaneously in other tracks. They can be dis-
played in curves with the identification
External 1...3
or used as variables
'Command name.EX1...3'
in formulae.
Open the dialog window
External measured value #
(see below) with the pa-
rameters for the external measured value.
Open the dialog window
External measured value #
(see below) to edit the
parameters of the external measured value.
Delete the selected external measured value.