Pump – Metrohm tiamo 1.1 Manual User Manual
Page 568

540
Method
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Target
Tower
[ 1 ], 2
Selection of the tower of the Sample Processor for moving to the required
lift position. Both Tower 1 and Tower 2 are always selectable even if the
Sample Processor has got only one tower.
Lift position
0 ... 235 mm, [ Work position ], Shift position, Rinse position, Special position,
Home position
(730, 778, 789, 814, 815, 855)
0 ... 100 mm, [ Work position ], Shift position, Rinse position, Special position,
Home position
(774 KF Oven Sample Processor)
Input the desired lift position in mm or selecting one of the predefined lift
positions
Work position, Shift position, Rinse position
or
Special position
. A lift
height of 0 mm corresponds to the
Home position
where the lift is moved to
the uppermost stop.
Note
Please make sure that none of the lift positions exceeds the maximum stroke
path given in the tower properties of the device.
Parameters
Lift rate
5 ... [ 25 ] °/s
(778, 789, 814, 815, 855)
3 ... [ 25 ] °/s
(730)
3 ... [ 12 ] °/s
(774)
Selection of the speed for moving the lift.
PUMP
Command for controlling pumps connected (or built-in) to the Sample Proces-
sor.
Devices
This command can be executed with the following devices:
Sample Processor: 730, 778, 789, 814, 815, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
PUMP
are configured in the following dialog
window:
•
PUMP