Metrohm tiamo 1.1 Manual User Manual
Page 56

28
General program functions
Identifier
Description
Commands
.SPO
Current external position (entry at end of command);
0 means invalid position
SWING
.STE
Start temperature (temperature after processing the
start conditions) in °C
DET, MET, SET, KFT,
KFC, STAT
.STY
Stop type with which the command has been stopped:
1 = normal;0 = manual or after error
DET, MET, SET, KFT,
STAT, MEAS, CAL
MEAS, STDADD, DOS
.SVA
Start volume absolute (volume added in accordance
with the start condition "Start volume") in mL
DET, MET, SET, KFT,
STAT
.SVM
Start volume measured value (volume added in accor-
dance with the start condition "Start measured value")
in mL
DET, MET
.SVS
Start volume measured value (volume added in accor-
dance with the start condition "Start slope") in mL
DET, MET
.SVT
Start volume total (volume added in accordance with
all three start conditions) in mL
DET, MET, SET, KFT,
STAT
.TITER
Titer value for the solution used for the command
DET, MET, SET, KFT,
STAT, STDADD dos,
STDADD auto, ADD,
DOS, LQH
.TOU
Timeout status: 1 = max. delay period elapsed;
0 = max. delay period not elapsed
RECEIVE, TRANSFER;
SCAN
.VAR
Variance of the calculated standard addition result in
selected unit
STDADD
.VOL
Added volume
STDADD, ADD, DOS,
LQH