Configuration
Settings according to 21 CFR 11
Sets the parameters to standard settings for FDA-conform work.
Inactivity delay
on, [ off ];
FDA-default:
on
1 ... [ 10 ] ... 60 min;
FDA-default:
10 min
If this option is switched on then the dialog window for signing will be
automatically closed when the entered time limit has expired.
Remove password after signature
on, [ off ];
FDA-default:
on
If this option is switched on then the password must be entered again af-
ter each signature.
Sign each determination separately
on, [ off ];
FDA-default:
on
If this option is switched on then each determination contained in the de-
termination overview must be signed individually.
Default reasons
On the tab
Default reasons
the reasons are defined that have to be entered when
methods and determinations have to be signed for, or when methods, determina-
tions and sample data are altered.
Category
[ Signature level 1 ], Signature level 2, Modification of methods, Modification of
determinations, Modification of sample data
Selection of the category for which the reasons are to be defined.
Reasons
50 characters
Shows the reasons defined for the selected category.
Move text upward (alters sequence).