Titer – Metrohm tiamo 1.1 Manual User Manual
Page 746

718
Configuration
Action
If during solution monitoring it is found that the working life has expired then one
of the following actions will be triggered automatically at Start test:
Record message
The message that the working life of the solution has expired will be auto-
matically saved with the determination.
Display message
A message will appear and you can choose whether to continue the run or
terminate it. If the run is continued then the message that the working life of
the solution has expired will be automatically saved with the determination.
Stop determination
The running determination will be automatically terminated. The following
message must be confirmed with
[OK]
.
Titer
Titer determination
Titer
Value:
1E-12 ...
[ 1.000 ] ... 1E+12, max. 10 characters
Unit:
10 characters, [ 'empty' ], mol/L, mmol/L, µmol/mL, g/L, mg/L, µg/mL,
mg/mL, ppm, %, mEq/L
Titer of the solution. The value for the titer of a solution used by the titration
commands
DET
,
MET
,
SET
and
KFT
is available as a variable
'Command
name.TITER'
for calculations.
Date titer det.
Date and time of last titer determination, which is automatically entered af-
ter each automatic titer determination or manual entry.
Titer method
Name of the method with which the last titer determination was carried out.
If the titer has been entered manually then
manual
will be shown here.
User
Short name of the user logged in during titer determination or when the titer
was entered manually. If work is not carried out using log in then the user
logged in under Windows will be entered automatically.