beautypg.com

34 advanced mode (user-defined), Working in advanced mode – Teledyne LeCroy Sierra M6-4 SAS_SATA Protocol Analyzer User Manual User Manual

Page 118

background image

Teledyne LeCroy

SAS/SATA Software Menus and Toolbars

116

 

 

Sierra M6‐4 SAS/SATA Protocol Analyzer User Manual

2.2.34

Advanced Mode (User-Defined)

The Advanced Mode expands Analysis capability by allowing you to program complex
triggering and data capture projects.

The Advanced Mode is a state machine with up to 23 different states. You can program
each state individually to:

Trigger on a different event or trigger unconditionally.

Capture Everything, Nothing, or a user‐defined pattern.

Include up to three ELSE IF statements, allowing a jump to any other state based
on a user definition.

Use up to three timers, which you can set to a maximum value of 42949 ms. You
can set a timer in the state or continue the timer set in the previous state.

Output an external trigger (High or Low).

Note:

In Advanced Mode, events on each link are counted independently. A condition is met if the

number of events on a link equals the defined occurrence.

Working in Advanced Mode

To start working in the Advanced Mode, click the Easy, Switch to Advanced Mode button
in an open Analyzer window.

You can:

Display the state definition

Set Output Trigger level

Select up to three timers

Define the If condition and up to three Else If conditions

Set number of occurrences before trigger

Set captured data

Set excluded data

Go to next state

Add state

Choose link for Sequencer setup