beautypg.com

Teledyne LeCroy Oscilloscope System Recovery User Manual

Oscilloscope system recovery, Important, Before you start

background image

Instructions

Oscilloscope System Recovery

Also:

Deutsch

,

Italiano

,

Français

Your oscilloscope is designed to operate reliably for many years.
The application software operating the instrument runs on a
Microsoft Windows

®

platform. The loading or incomplete

removal of additional Windows applications can cause problems
in the stability of the operating system. Severe cases may require
you to reload the base operating system and oscilloscope
application.

For this purpose, Teledyne LeCroy provides an Acronis True
Image

®

recovery application and a backup image in an extra

partition on the instrument's hard drive. The recovery process is
easy to perform.

Since third-party recovery software is used, our instructions may
not be the most definitive or current. We encourage you to take
advantage of additional resources available from the Acronis
website at:

www.acronis.com/homecomputing/download/docs/

Important!

Depending on the year and model, your oscilloscope may be
installed with either True Image Home or True Image Echo
Workstation. It is critical to check which product you are using,
as following the incorrect procedure can cause the system to fail.

You must reactivate Windows after the restoration is complete,
either by Internet connection to Microsoft's website or by
telephone. Have your Windows Product Key number handy for
Widows reactivation.

Before You Start

l

Find your Windows Product Key number, usually listed on a
sticker on the back of the oscilloscope.

NOTE: In LabMaster models, the Windows Product Key is
found on the removable hard drive. Locate the drive and
power down the oscilloscope before removing the hard drive
to find the product key.
Replace the drive before powering
on again at the start of the recovery procedure.

l

If you intend to reactivate Windows through the Internet,
connect a network cable to the oscilloscope's LAN port.

l

Connect a keyboard and mouse via the USB host ports.

Determine the Acronis Product

1. Apply power to the oscilloscope.

2. During the startup process, as soon as you see the message

"Starting Acronis Loader... Press F11 for Acronis Startup
Recovery Manager," press the F11 key until the Acronis logo
appears momentarily. The Acronis window is displayed.

NOTE: Do not press F11 before you see the "Starting Acronis
Loader..." message or you will enter the boot device
selection menu. If a bootmenu dialog box appears, press
Cancel or Esc.

3. Check the Acronis window to determine which product you

are using. Follow the steps for Using Home or Using Echo
Workstation accordingly.

Using Home

1. Select Acronis True Image Home (Full Version).

2. On the Acronis True Image Home page, under options for

Recover, select My Disks. The Recovery Wizard opens.

3. On the Recovery Wizard, under Archive Selection, select the

disk archive that has a create date, then click Next.

4. Under Recovery Method, select Recover whole disks and

partitions, then click Next.

5. Under What to Recover, select NTFS (SYSTEM) (C:), then

click Next.

6. Under Settings of Partition C, in the top section, Partition

location (required), select New Location.The Partition
Destination window opens.

7. Under New Partition Location, select NTFS (SYSTEM) (D:),

then click Accept. This returns you to the Settings of
Partition C step. Click Next.

8. A summary window is displayed indicating that Acronis

True Image is ready to proceed with recovering partition C -
> D. Click Proceed to start the recovery process.

NOTE: Recovery takes approximately 4 to 15 minutes
depending on the version of operating system that is being
restored. The progress is displayed on the screen.

9. When recovery is complete, you will see the message

"Recover operation succeeded." Click OK.

10. Click the close button to exit the Acronis window. The

oscilloscope will restart and begin installing the required
software. Continue to reinstall the oscilloscope application.

NOTE: If prompted to restart Windows, select Restart Later.

Using Echo Workstation

1. On the Pick a Task page, select Recovery. The Restore Data

Wizard opens.

2. Click Next on the Welcome page.

3. On Backup Archive Selection, choose Acronis Secure Zone

(this is where the recovery data is located on your
oscilloscope). Click Next.

4. On Restoration Type Selection, select Restore disks or

partitions, then click Next.

5. On Partition or Disk to Restore, select SYSTEM (C:) as the

source, then click Next.

6. On Restored Partition Location, select SYSTEM (C:), then

click Next.

7. On Restored Partition Type, select Active, then click Next.

Note: If a window appears asking if you want to buy Acronis
products, check "Do not show this message again", then
click OK.

8. On Restored Partition Size, simply click Next.

9. On Next Selection, select No, I do not (want to restore

another partition), then click Next.

10. On Restoration Options, simply click Next to use the default

923605 Rev B

1