Manual control

71

General

Information about the attached rack is shown here. In addition the speed of rota-
tion and the lift speed for manual sample changer control can be set here.

Rack name

Shows the name of the attached rack. If no rack is attached then

--------

will

be shown.

Rack code

Shows the rack code of the attached rack. If no rack is attached then

-------

-

will be shown.

Number of positions

Shows the number of positions of the attached rack. If no rack is attached
then

--------

will be shown.

Shift rate

5 ... [ 20 ] °/s

Speed of rotation for manual operation of the sample changer.

Lift rate

3 ... [ 25 ] mm/s

Lift speed for manual operation at the selected tower.

Swing rate

10 ... [ 55 ] °/s

Swing speed for manual operation of the swing head on the sample
changer.

Initialize the attached rack. During initialization the cursor appears as an
hourglass.

Note
During the initialization of the rack the following actions are carried out:
– Rack rotates to position for reading the rack code.

– Rack data from rack code is transferred to sample changer.
– Lifts are raised to 0 mm.
– Swing head moves back.