Metrohm tiamo 1.0 Manual User Manual
Page 637

Subwindow Devices
613
Lift positions
Tower 1
Defines the lift positions for Tower 1. These apply for all rack positions ex-
cept those that are defined as Special beaker.
Work position
0 ... [ 83 ] ... 100 mm
Work position for Lift 1. At this lift position the electrodes, stirrer and buret
tips are optimally positioned for work.
Rinse position
[ 0 ] ... 100 mm
Rinse position for Lift 1. This lift position is used for rinsing the electrodes.
Shift position
[ 0 ] ... 100 mm
Shift position for Lift 1. Each time that the rack rotates the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here then the rotation will take place at
the current lift position. This means that the shift position must be selected
so that a safe rotation over the whole rack is possible at any time.
Special position
[ 0 ] ... 100 mm
Special position for Lift 1. This additional definable position can be used
e.g. during pipetting so that the tip is just immersed in the sample solution.
Special beakers
Special beaker
Shows the number of the special beaker for the selected rack.
Rack position
Shows the number of the rack position for the special beaker.