Commands

331

External measured value

With the option

Additional measured value

3 new measured values can be taken

over from existing measuring commands (i.e.

MEAS

). They can be displayed in

curves with the identification

External 1...3

or used as variables

'Command

name.EX1...3'

in formulae.

Command

Selection from existing measuring commands

Selection of the command from which the additional measured value will be
taken over.

Quantity

Selection from existing measuring quantities

Selection of the quantity from the command which will be adopted as quan-

tity of the additional external measured value.

Unit

Unit of calculated measured value.

Comment

250 characters, [ 'empty' ]

Comment about the external measured value.

DET U

Command for Dynamic Equivalence Point Titrations with potentiometric volt-
age measurement.

Devices

This command can be executed with the following devices which have the DET
mode:

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Titrando: 808, 809, 835, 836

Appearance

The command has the following appearance:

Parameters

The parameters for the command

DET U

are configured in the following 7 tabs:

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.