Metrohm tiamo 1.0 Manual User Manual
Page 528

504
Method
Command name
25 characters, [ Data request # ]
Name of the command.
Sample data request
The sample data to be entered only while the determination is running can
be selected here. The dialog window
Sample data request
with the sample
parameters defined here will appear during the method run. The sequence
is halted until the input is confirmed.
Note
In case the
sequence should not be stopped (for example for back-weighing
in KF titrations), the REQUEST command has to be inserted in another track than
the one with the titration command. The option
Return immediately
must be acti-
vated in this track.
Sample position
on, [ off ]
Switching on/off the request for the sample position.
ID1 ... ID8
on, [ off ]
Switching on/off the request for the sample identifications
ID1 ... ID8
. If
other names were given for this variables in the
START
command, these are
displayed in the dialog window
Sample data request
.
Note
The sample identifications
ID1 ... ID8
are by default of the type
Text
. If you wish
to enter numbers for this method variables for using them in later calculations,
you have to change the type to
Number
in the
START
command.
Sample size
on, [ off ]
Switching on/off the request for the
Sample size
.
Unit
on, [ off ]
Switching on/off the request for the
Sample size unit
.
Message
on, [ off ]; Text (unlimited)
This message is displayed in the dialog window
Sample data request
. The
Text editor to enter or change the message is opened with
or by
double-clicking into the message field.
Note
In case the option
Comments on modification of sample data (live)
is activated in
the Security settings, the dialog window
Modification comment sample data
(see
Subwindow Run - Single determination - Modification comment sample data) is
displayed after entering the sample data. A
Reason
and a
Comment
must be en-
tered for the modification.