Metrohm tiamo 1.0 Manual User Manual
Page 419

Commands
395
Switch off automatically
[ on ], off
(only Titrando)
If this option is enabled, the stirrer will be switched off automatically when
the command is finished.
Switch on/off automatically
[ on ], off
only Titrino 751, 785, 798, 799)
If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command.
Start conditions
Screenshot and parameters: see SET pH - Start conditions
Control parameters
Command name
25 characters, [ KFT Ipol # ]
Name of the command.
Endpoint
EP at
-2000.0 ... [ 250.0 ] ... 2000.0 mV, off
Measured value for the endpoint.
Titration rate
Titration rate
Three predefined sets of parameters can be selected for the titration rate:
slow
,
optimal
and
fast
. For these sets the parameters for the
Control
are not
displayed. In order to edit these parameters the value
user
must be se-
lected.
slow
For titrations in which the finest details are also to be visible. However, this
can also increase noise and lead to the production of unwanted EPs.
[ optimal ]
Parameter set for all standard titrations; optimized for the most frequent
applications.