Formula editor

21

Identifier

Description

Commands

.LCO

Loop counter = current number of completed loops both for

Repeat and While loops

LOOP, CAL LOOP

.LP.CA#

Calculated value # (1...3) for the last measuring point

DET, MET, SET,
KFT, KFC ,MEAS

.LP.CHA

Charge for last measuring point in mA·s

KFC

.LP.DME

Difference in measured values for last measuring point

MET

.LP.DVT

dV/dt for the last measuring point (SET) or drift for last measuring
point in µg/min (KFC)

SET, KFC

.LP.ERC

ERC for last measuring point

DET

.LP.EX#

External value # (1...3) for last measuring point

DET, MET, SET,
KFT, KFC, MEAS

.LP.GFL

Gas flow value for last measuring point in mL/min

MEAS T/Flow

.LP.IGE

Current pulse current for last measuring point in mA

KFC

.LP.QTY

Measured value (water) for last measuring point in µg

KFC

.LP.TEM

Temperature for last measuring point in °C

DET, MET, SET,
KFT, MEAS (without

T and T/Flow), CAL
MEAS

.LP.TIM

Time for reaching last measuring point in s

DET, MET, SET,
KFT, KFC, MEAS,

CAL MEAS

.LP.UGE

Potential at generator electrode for last measuring point; 0 = not
defined
, 1 = <14 V, 2 = 14...28 V, 3 = >28 V

KFC

.LP.VOL

Volume for last measuring point in mL

DET, MET, SET, KFT

.LP.MEA

Measured value for last measuring point in the units of the

measured value

DET, MET, SET,

KFT, KFC, MEAS,
CAL MEAS

.LPO

Current absolute lift position in mm (entry at end of command)

LIFT

.LST

Start time of loop command (

Date/Time

)

LOOP, CAL LOOP

.MA.GFL

Maximum gas flow in mL/min

MEAS T/Flow

.MA.MEA

Maximum measured value (U

ind

) in mV

MEAS

.MA.MEA

Maximum measured value in the units of the measured value

DET, MET, SET, KFT

.MA.TEM

Temperature for maximum measured value in °C

DET, MET, SET,
KFT, MEAS (without
T and T/Flow)

.MA.TIM

Time for reaching the maximum measured value in s

DET, MET, SET,

KFT, MEAS

.MA.VOL

Volume for maximum measured value in mL

DET, MET, SET, KFT

.MI.GFL

Minimum gas flow in mL/min

MEAS T/Flow

.MI.MEA

Minimum measured value (U

ind

) in mV

MEAS

.MI.MEA

Minimum measured value in the units of the measured value

DET, MET, SET, KFT

.MI.TEM

Temperature for minimum measured value in °C

DET, MET, SET,

KFT, MEAS (without