Configuration data
551
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
Example:
************1*
sets output line 1 to active (= set) which, e.g.,
would be a stop command for a connected Titrino.
************0*
sets
the line to inactive.
Note
We recommend that all the non-relevant output lines are masked with an asterisk
*
so that the conditions of these lines are not altered.
Pulse length
100 ... [ 200 ] ... 1000 ms
Duration of the transmitted pulse.
Predefined templates
In order to simplify the use of remote control commands, particularly when several
devices are connected together with Metrohm cables, the following predefined bit-
patterns are available as command parameters:
Parameter
Bit-pattern
Function
Start Dos1
*******p******
Starts Dosimat 1 (Titrino via "activate")
Start Dos2
*****p********
Starts Dosimat 2 ( " )
Start Dos*
*****p*p******
Starts Dosimats 1 and 2 ( " )
6.3.4 Options
Overview
Under
Tools, Options
general program properties can be set on the following 3
tabs:
•
General
Selection of the dialog language and switches the emergency stop
button on/off.
•
Save
Saves settings on exiting tiamo.
•