beautypg.com

Metrohm tiamo 1.0 Manual User Manual

Page 356

background image

332

Method

Start conditions

Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-

tion.

Titration parameters

Parameters for the run of the titration.

Stop conditions

Definitions of conditions which cause the stop of the titration.

Potentiometric evaluation

Parameters for the potentiometric evaluation of titration curves.

Additional evaluations

Definition of additional methods for evaluation of the titrations curves.

Additional measured values

Definition of additional measured values of other measuring com-
mands which are saved as additional column in the list of measuring

points.

General/Hardware

Command name

25 characters, [ DET U # ]

Name of the command.

The general parameters for the control device, the dosing device, sensor and the

stirrer are defined on this tab.

Device

Device name

Device name, [ not defined ]

Selection of the device from device table. Only those devices being able to
execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the

start of the method.