Determination run – Metrohm tiamo 1.0 Manual User Manual
Page 145

Subwindow Run
121
Dark orange background
Selected currently running line. If sample data is only altered after the start
of the determination (in the status
BUSY
) then this is regarded as being a
live modification.
White background
Lines that have not yet been processed. These lines can be edited.
Turquoise background
Selected line that has not yet been processed.
Loaded
Shows the sample table whose data has been loaded into the working
sample table with
Sample table, Load...
. If no sample table has been loaded
or if all the lines have been deleted then this field will be empty. If data in a
loaded sample table is later altered or if new lines are added then in addi-
tion to the name
(modified)
will be shown.
For possible ways of editing the working sample table and the window menu
please refer to Edit working sample table.
Determination run
In a series of determinations that are started on the tab
Determination series
in the
subwindow
Run
each determination takes place as follows:
1 - Load sample data
The sample data entered in the working sample table for the current line will be
loaded for the determination.
2 - Select method
Method selection triggers the following actions:
• Select method without assignment ID
If the option
Method selection by assignment ID
is not switched on the
tab
Process
then the method selected in the field
Method
will remain
loaded.
• Select method with assignment ID
If the option
Method selection by assignment ID
is switched on the tab
Process
then the
Sample assignment ID
will be used to load the method
defined in the
Sample assignment table
.
3 - Carry out start test
During the Start test the following checks and actions are triggered:
• Right to carry out the method?
Checks whether the logged-in user has the right to execute the par-
ticular method.
• Method executable?
Checks whether the particular method is executable. If the method is
non-executable then a message appears asking for the method test to
be carried out and the method edited accordingly.
• Check device (module) and rack
Checks whether the devices (or device types), modules (e.g. measur-
ing inputs, towers, etc.) and racks demanded in the device-specific
commands are present and available or assigned. If the device has
not yet been assigned in a command then the window
Device assign-
ment
appears for assigning the device.