General program functions
System variables
System variables are general variables that are not assigned to individual com-
mands or determinations that are accepted for the determination at the start of the
determination. In the
Variables
field of the Formula editor all the
System variables
that are available for the current method are listed.
Syntax
'SV.Variable identifier'
Examples:
'SV.SIN', 'SV.SLI'
In order to avoid syntax errors the system variables can be selected directly
in the Formula editor under
Variables/System variables
.
System variables
Identifier Description
.ACC
Autostart current counter (
Number
)
.ACE
Autostart set counter (
Number
)
.FUN
Full name of logged-in user (
Text
)
.REM
Remarks (
Text
)
.RUN
Sample number (
Number
)
.SEN
Indicates whether the end of the sample table has been reached; 1 = yes, 0 = no (
Num-
ber
)
.SIN
Indicates whether the determination was started as a single determination or within a se-
ries; 1 = single determination, 0 = series determination (
Number
)
.SLI
Sample table current line (
Number
)
.STA
Indicates whether statistics are switched on; 1 = yes, 0 = no (
Number
)
.STC
Start counter (
Number
)
.USN
Short name of logged-in user (
Text
)
.ORG
Method sequence: 1 = original determination, 0 = reprocessed (
Number
)
.STO
Indicates whether the determination was stopped (manual stop, stop via a SEND com-
mand, emergency stop); 1 = stopped, 0 = terminated normally (
Number
)