671

higher lift position than that defined here then the rotation will take place at
the current lift position. This means that the shift position must be selected

so that a safe rotation over the whole rack is possible at any time.

Special position

[ 0 ] ... 235 mm

Special position for Lift 1. This additional definable position can be used
e.g. during pipetting so that the tip is just immersed in the sample solution.

Tower 2

Defines the lift positions for Tower 2. These apply for all rack positions ex-

cept those that are defined as Special beaker.

Work position

[ 0 ] ... 235 mm

Work position for Lift 2. At this lift position the electrodes, stirrer and buret
tips are optimally positioned for work.

Rinse position

[ 0 ] ... 235 mm

Rinse position for Lift 2. This lift position is used for rinsing the electrodes.

Shift position

[ 0 ] ... 235 mm

Shift position for Lift 2. Each time that the rack rotates the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here then the rotation will take place at

the current lift position. This means that the shift position must be selected
so that a safe rotation over the whole rack is possible at any time.

Special position

[ 0 ] ... 235 mm

Special position for Lift 2. This additional definable position can be used
e.g. during pipetting so that the tip is just immersed in the sample solution.

Special beakers