beautypg.com

Metrohm tiamo 1.0 Manual User Manual

Page 643

background image

Subwindow Devices

619

Stop bit

[ 1 ], 2

Number of stop bits.

Handshake

[ None ], HW, SW

Type of data transmission protocol.

Timeout

1000 ... [ 2000 ] ... 9000 ms

Maximum waiting time for receiving characters. If this time is exceeded the
reception will be stopped.

Terminator for send

\0D, \0D\0A, \0D\0A\0A, \0D\0D\0A, [ 'empty' ]

Final character for transmission in hexadecimal code.

Terminator for receive

\0D, \0D\0A, \0D\0A\0A, \0D\0D\0A, [ 'leer' ]

Final character for reception in hexadecimal code.

Code page

[ Cp437 ], Cp850, Cp851

Code page used for the data transmission.

Tests the connection to the balance. The dialog window

Check connection

appears.

Test connection

In order to test the connection to a balance the balance must be selected in the
Device table and then under

Edit, Properties

the Properties window for this device

opened. On the tab

RS 232

the button

[Check connection]

must be pressed.

Opens the dialog window

Check connection

.

If the print key on the connected balance is pressed then the data provided
by the balance will be inserted in the following fields:

Device name

Name of balance.

Sample size

Sample weight.

Sample size unit

Sample weight unit.

ID1

Sample identification ID1.

ID2

Sample identification ID2.