Metrohm tiamo 1.0 Manual User Manual
Page 473

Commands
449
CAL LOOP pH - properties
The number of buffers and the buffer type are defined in the dialog window
CAL
LOOP pH
.
Command name
25 characters, [ Calibration loop pH # ]
Name of the command.
Buffers
Number of buffers
1 ... [ 2 ] ... 5
Selecting the number of buffers for the pH calibration.
Buffer type
Selecting a predefined buffer series for the calibration.
[ Metrohm ], NIST, DIN, Fisher, Fluka Basel, Mettler, Merck, Beckman, Radiome-
ter
If these buffers are used, tiamo recognizes the particular buffer automati-
cally.
Custom
An own buffer series defined in the configuration under
Tools, Templates,
Custom calibration buffers
can be used with
Custom
.
Special
If
Special
is selected then up to five calibration buffers can be defined di-
rectly in the dialog window of the command
CAL LOOP pH
.
Note
The automatic buffer recognition is inactive for the buffer type
Special
. The buf-
fers must therefore be used in the same order as they have been defined in the
command
CAL LOOP pH
.
Buffer # pH
-20.000 ... 20.000
pH value of the special buffers. This parameter is only displayed for
Buffer
type
=
Special
.
Note
Please note that you must enter the pH values corresponding to the measuring
temperature used. During the method run a message is displayed to indicate the
next buffer to be measured.