Metrohm tiamo 1.0 Manual User Manual
Page 43

Formula editor
19
Identifier
Description
Commands
.BLV
Blank value of the sensor used for the command (for ISE sensors
only) or the blank value calculated during calibration (for CAL
MEAS Conc)
DET U, MET U, SET
U, MEAS U, MEAS
Conc, CAL MEAS
Conc
.BP{#}.DME
Difference in measured values for break point # (1...9)
MET
.BP{#}.ERC
ERC or 1st derivative for break point # (1...9)
DET
.BP{#}.MEA
Measured value for break point # (1...9) in the units of the meas-
ured value
DET, MET, MEAS
(without T and
T/Flow)
.BP{#}.TEM
Temperature for break point # (1...9) in °C
DET, MET, MEAS
(without T and
T/Flow)
.BP{#}.TIM
Time for breal point # (1...9) in s
DET, MET, MEAS
(without T and
T/Flow)
.BP{#}.VOL
Volume at break point # (1...9) in mL
DET, MET
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 = READY;
invalid (variable not present) = Command has never been
started
All
.CBY
Command status; 1 = conditioning active, 0 = conditioning
not active
SET, KFT, KFC
.COK
Command status; 1 = conditioning criterion fulfilled, 0 =
conditioning criterion not fulfilled
SET, KFT, KFC
.CONC
Concentration of the solution used for the command
DET, MET, SET,
KFT, ADD, LQH
.DBL
Total duration for processing the command in s
DET, MET, SET,
KFT, KFC, MEAS,
CAL MEAS
.DRI
Current or last drift for drift correction in µL/min
SET, KFT, KFC
.DCS
Duration for processing all start conditions in s
DET, MET, SET, KFT
.DTI
Time for the drift correction (time from start of titration until end of
command) in s
SET, KFT, KFC
.EGF
Last measured gas flow (measured value after processing the
command) in mL/min
MEAS T/FLow
.EME
Final measured value (measured value after processing the
command) in the units of the measured value
DET, MET, SET,
KFT, KFC, MEAS,
CAL MEAS
.ENP
Electrode zero point of the sensor used for the command (di-
mensionless for pH sensor or in mV for ISE sensor) or the elec-
trode zero point calculated from the calibration (for CAL LOOP)
DET, MET, SET,
KFT, MEAS, CAL
LOOP
.EP{#}.DME
Difference in measured values for end point # (1...9)
MET
.EP{#}.DVT
Drift for end point # (1...9) in µg/min
KFC
.EP{#}.ERC
ERC for the end point # (1...9)
DET
.EP.MEA
Measured value for the end point in mV
KFC
.EP{#}.MEA
Measured value for the end point # (1...9) in the units of the
measured value
DET, MET, SET, KFT