beautypg.com

Metrohm tiamo 1.0 Manual User Manual

Page 43

background image

Formula editor

19

Identifier

Description

Commands

.BLV

Blank value of the sensor used for the command (for ISE sensors

only) or the blank value calculated during calibration (for CAL
MEAS Conc)

DET U, MET U, SET

U, MEAS U, MEAS
Conc, CAL MEAS
Conc

.BP{#}.DME

Difference in measured values for break point # (1...9)

MET

.BP{#}.ERC

ERC or 1st derivative for break point # (1...9)

DET

.BP{#}.MEA

Measured value for break point # (1...9) in the units of the meas-

ured value

DET, MET, MEAS
(without T and
T/Flow)

.BP{#}.TEM

Temperature for break point # (1...9) in °C

DET, MET, MEAS
(without T and

T/Flow)

.BP{#}.TIM

Time for breal point # (1...9) in s

DET, MET, MEAS
(without T and
T/Flow)

.BP{#}.VOL

Volume at break point # (1...9) in mL

DET, MET

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 = READY;
invalid (variable not present) = Command has never been
started

All

.CBY

Command status; 1 = conditioning active, 0 = conditioning
not active

SET, KFT, KFC

.COK

Command status; 1 = conditioning criterion fulfilled, 0 =
conditioning criterion not fulfilled

SET, KFT, KFC

.CONC

Concentration of the solution used for the command

DET, MET, SET,
KFT, ADD, LQH

.DBL

Total duration for processing the command in s

DET, MET, SET,

KFT, KFC, MEAS,
CAL MEAS

.DRI

Current or last drift for drift correction in µL/min

SET, KFT, KFC

.DCS

Duration for processing all start conditions in s

DET, MET, SET, KFT

.DTI

Time for the drift correction (time from start of titration until end of

command) in s

SET, KFT, KFC

.EGF

Last measured gas flow (measured value after processing the
command) in mL/min

MEAS T/FLow

.EME

Final measured value (measured value after processing the

command) in the units of the measured value

DET, MET, SET,

KFT, KFC, MEAS,
CAL MEAS

.ENP

Electrode zero point of the sensor used for the command (di-
mensionless for pH sensor or in mV for ISE sensor) or the elec-
trode zero point calculated from the calibration (for CAL LOOP)

DET, MET, SET,
KFT, MEAS, CAL
LOOP

.EP{#}.DME

Difference in measured values for end point # (1...9)

MET

.EP{#}.DVT

Drift for end point # (1...9) in µg/min

KFC

.EP{#}.ERC

ERC for the end point # (1...9)

DET

.EP.MEA

Measured value for the end point in mV

KFC

.EP{#}.MEA

Measured value for the end point # (1...9) in the units of the

measured value

DET, MET, SET, KFT