Flow – Metrohm tiamo 1.0 Manual User Manual
Page 501

Commands
477
causes heating at the maximum possible heating rate (see Instructions for
Use for 774 Oven Sample Processor).
Timeout
1 ... 999 min, [ off ]
Should the target temperature not be reached within the defined
Heating
time
this waiting time starts. This parameter is only visible if a
Heating time
has been defined.
Action if timeout expires
on, [ off ]
(displayed only if
Timeout
≠
off
)
Selection of the action to be carried out if the target temperature is not
reached after the expiry of the timeout.
Display message
[ on ], off
A message is displayed and you can choose whether you want to continue
the run or cancel it.
Stop determination
on, [ off ]
The determination is stopped and the next determination of the series is
started.
Stop determination and series
on, [ off ]
The determination as well as the series are stopped. An Error track will be
started if there is one defined.
FLOW
Command for controlling the gas flow of the 774 Oven Sample Processor.
Devices
This command can only be executed with the 774 Oven Sample Processor.
Appearance
The command has the following appearance:
Parameters
The parameters for the command
FLOW
are configured in the following dialog
window:
•
FLOW