Commands
359
Upper limit
-200.0 ... [ 200.0 ] µA
(for
Window
=
Measured value
)
0.00000 ... [ 9999.99 ] mL
(for
Window
=
Volume
)
0 ... [ 999999 ] s
(for
Window
=
Time)
Value for the upper limit of the window.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
MET
Command for Monotonic Equivalence Point Titrations (
MET
).
Principle
With this command titrations with reagent addition at constant volume increments
are carried out. Measured value acceptance is drift-controlled (equilibrium titra-
tion) or after a waiting period. Equivalence points are evaluated automatically.
Note
This command is suited for titrations with relatively high signal variations or sud-
denly occurring potential jump and for slow titrations or slowly responding e-
lectrodes.
Commands
Depending on the measured value one of the following four
MET
commands can
be chosen:
• MET pH
Potentiometric pH measurement with pH electrodes (measured quan-
tity pH).
• MET U
Potentiometric voltage measurement with metal electrodes (measured
quantity voltage U).
• MET Ipol
Voltametric measurement with selectable polarization current (meas-
ured quantity voltage U).