Metrohm tiamo 1.0 Manual User Manual
Page 484

460
Method
Command name
25 characters, [ Liquid Handling # ]
Name of the command.
The general parameters for the control device and the dosing device are defined
on this tab.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Dosing device
Dosing device
[ 1 ] ... 4
(Titrando, 846)
[ 1 ] ... 3
(778, 789, 814, 815)
Selection of the dosing device to be used for the liquid handling tasks. All
the dosing device connections (MSB) are displayed which are possible with
the selected device type.
Solution
Titrant/Solution, [ not defined ]
Selection of the titrant or solution defined in the solution table.
Parameters
Command name
25 characters, [ Liquid Handling # ]
Name of the command.
The liquid handling parameters are defined on this tab.
Function
Selecting the liquid handling function to be carried out: