With the context-sensitive menu item

Live modifications

the dialog window

Live

modifications

opens, in which sample data can be altered live during a running de-

termination.

Sample position

[ 1 ] ... 999

Position of the sample on the rack. This number is used to move to the
sample position with the command

MOVE

. This means that

Target

=

Sample

position

must be set.

ID1 ... ID8

50 characters, selection from text templates

Sample identifications ID1...ID8. In this field any text can be entered, a

check for type and limits only takes place at the start of the determination.
For sample identifications of the type

Date/Tim

e the entry must be made in

the format

YYYY-MM-DD hh:mm:ss

. If Text templates have been defined for

the sample identifications then these can also be selected.

Sample size

[ 1.0 ], numerical with max. 10 numbers

Sample size (weight). In this field you can only enter numbers, decimal
point,

+

, –,

E

and

e

. The limits will only be checked at the start of the deter-

mination.

Sample size unit

16 characters, [ g ], mg, µg, mL, µL, pieces

Sample size unit (weight). Any text can be entered in this field.

Note

In principle it is not possible to enter a formula in the

Sample position

,

ID1...ID8

,

Sample size

and

Sample size

unit

fields.

If the option

Comment on modification of sample data

is switched on in the Securi-

ty settings window then before saving the window

Modification comment sample

data

opens.

Modification comment for sample data

If the option

Comments on modification of sample data

is switched on in the Security

settings then before the edited sample data is accepted the window

Modification

comment sample data

opens first, in which a reason for the modification must be

selected and comments about the modification must be entered.