Metrohm tiamo 1.0 Manual User Manual
Page 687

Subwindow Common Variables
663
Stop determination
The running determination will be automatically terminated. The following
message must be confirmed with
[OK]
.
History
History table
The table contains the last 10 values for the selected common variable and can-
not be edited or sorted. The individual values are sorted according to date so that
the most up-to-date assignment is shown last.
Value
The values are entered in the following colors:
- Blue, if the value has been assigned automatically by a method.
- Black, if the value has been entered manually.
- Orange, if the value is outside the warning limits.
- Red, if the value is outside the action limits
If the warning or action limits are infringed then the line number will also be
shown with an orange or red background respectively.
Assignment date
Date and time of value assignment.
Assignment method
Name of the method with which the value assignment was carried out. If the
value was entered manually then
manual
will be shown here.
User
Short name of the user who was logged in during value assignment or who
entered the value manually. If work is not carried out under login then the
name of the user logged in under Windows will automatically be entered
here.
History graph
The graph shows the last 10 values for the selected common variable. As in the
history table, the values are shown in different colors: